The water meter test bench is a device used for testing and calibrating the measurement accuracy of water meters, widely used by water utilities and metrology institutions. It simulates various flow conditions to ensure that the water meter meets national accuracy standards under different operating scenarios.
·Verification methods include start-stop method, gravimetric method, and diverter method.
·The verification process is automated, featuring a visual interface for control, automatic data storage, and report generation.
·Customization services are available.
·Capable of testing various types of mechanical water meters and ultrasonic water meters.
Model
|
SHA25-1
|
SHA25-P-1
|
SHA50-1
|
SHA50-P-1
|
Flow Range
|
(0.015-0.8) m³/h
|
(0.002-0.8) m³/h
|
(0.015-32) m³/h
|
(0.002-32) m³/h
|
Piston Specifification
|
/ |
Volume 22L/resolution 0.036ml/
pressure PN16/Panasonic servo drive
|
/ |
Volume 22L/resolution 0.036ml/
pressure PN16/Panasonic servo drive
|
Master Standard
|
Electronic scale
|
Electronic weight + piston
|
Electronic scale
|
Electronic weight + piston
|
Master Standard Specifification
|
Toledo 120kg/6000E
|
Toledo 60kg(6000E)/600kg(6000E)
|
||
Weighing Capacity
|
120L
|
60L/600L
|
||
Water Pump
|
Grundfos2.2KW(other brands also available)
|
Grundfos7.5KW (other brands are also available)
|
||
Commutator
|
DN25
|
DN25+DN50
|
||
Test Quantity
|
DN15*10 DN20*10 DN25*8 (Detection quantity
and dual lines can be customized)
|
DN15*10 DN20*8 DN25*6 DN32*5 DN40*4 DN50*3
(Detection quantity and dual lines can be customized)
|
||
Dimension
|
(5800 mm*1500 mm*2300mm)
|
(5800 mm*2000 mm*2700mm)
|
||
Electromagnetic Flowmeter
|
Japan Yokogawa AXG/DN2.5+DN25(other brands available)
|
Japan Yokogawa AXG/DN2.5+DN25+DN50(other brands available)
|
||
Test Method
|
Start-stop method + commutator method
|
|||
Uncertainty
|
≤0.2%(k=2)
|
|||
Pressure Range
|
0-16bar
|
|||
Ambient Temperature
|
15-30ºC
|
|||
Relative Temperature
|
45%-75%
|
|||
Atmospheric Pressure
|
(86-106)kPa
|
|||
Electricity and Water
|
380V, 50Hz/60Hz, 3-phase 5-wire power supply/pure water
|
|||
Air Source Pressure
|
(3.5-6)bar
|
|||
Data Collection
|
MBUS/RS485/ infrared/pulse/image acquisition, etc
|
|||
Temperature Sensor
|
PT100, accuracy level A; Installed at import and export
|
|||
Pressure Sensor
|
Pressure transmitter, accuracy 0.5%; Installed at import and export
|