Customization: | Available |
---|---|
After-sales Service: | Free After-Sales Support |
Warranty: | 1 Year |
Still deciding? Get samples of $ !
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
The fully automatic water meter testing bench is an automated device used for the verification, calibration, and performance testing of water meters. It is primarily utilized by water meter manufacturing enterprises, metrology inspection agencies, and water supply companies. This equipment can conduct flow tests and error verification on water meters of various diameters.
Improve testing efficiency: fully automated process with reduced manual intervention;
Data traceability: automatically generate test reports;
Strong scalability: customizable number of test positions and meter sizes according to customer requirements.
Model
|
SHA25-1
|
SHA25-P-1
|
SHA50-1
|
SHA50-P-1
|
Flow Range
|
(0.015-0.8) m³/h
|
(0.002-0.8) m³/h
|
(0.015-32) m³/h
|
(0.002-32) m³/h
|
Piston Specifification
|
/ |
Volume 22L/resolution 0.036ml/
pressure PN16/Panasonic servo drive
|
/ |
Volume 22L/resolution 0.036ml/
pressure PN16/Panasonic servo drive
|
Master Standard
|
Electronic scale
|
Electronic weight + piston
|
Electronic scale
|
Electronic weight + piston
|
Master Standard Specifification
|
Toledo 120kg/6000E
|
Toledo 60kg(6000E)/600kg(6000E)
|
||
Weighing Capacity
|
120L
|
60L/600L
|
||
Water Pump
|
Grundfos2.2KW(other brands also available)
|
Grundfos7.5KW (other brands are also available)
|
||
Commutator
|
DN25
|
DN25+DN50
|
||
Test Quantity
|
DN15*10 DN20*10 DN25*8 (Detection quantity
and dual lines can be customized)
|
DN15*10 DN20*8 DN25*6 DN32*5 DN40*4 DN50*3
(Detection quantity and dual lines can be customized)
|
||
Dimension
|
(5800 mm*1500 mm*2300mm)
|
(5800 mm*2000 mm*2700mm)
|
||
Electromagnetic Flowmeter
|
Japan Yokogawa AXG/DN2.5+DN25(other brands available)
|
Japan Yokogawa AXG/DN2.5+DN25+DN50(other brands available)
|
||
Test Method
|
Start-stop method + commutator method
|
|||
Uncertainty
|
≤0.2%(k=2)
|
|||
Pressure Range
|
0-16bar
|
|||
Ambient Temperature
|
15-30ºC
|
|||
Relative Temperature
|
45%-75%
|
|||
Atmospheric Pressure
|
(86-106)kPa
|
|||
Electricity and Water
|
380V, 50Hz/60Hz, 3-phase 5-wire power supply/pure water
|
|||
Air Source Pressure
|
(3.5-6)bar
|
|||
Data Collection
|
MBUS/RS485/ infrared/pulse/image acquisition, etc
|
|||
Temperature Sensor
|
PT100, accuracy level A; Installed at import and export
|
|||
Pressure Sensor
|
Pressure transmitter, accuracy 0.5%; Installed at import and export
|